OVERRIDE
MIRA\OVERRIDE TEM[PLOAD] | EFF[ICIENCY] | GAIN[IMAGE] value1 value2 |
TAMB[PID] value1 value2 [/FEBE ifb|all] [/RESET]
OVERRIDE allows to change parameters relevant for calibration. By default,
the parameters for frontend-backend combination 1 are changed (see below
for option /FEBE). Instead of value1 and value2, asterisks may be used to
leave the parameter unchanged (see below). The parameters changed are
displayed on the screen. If the scan loaded is a calibration (i.e. sky,
hot load, cold load), the calibration parameters are automatically
re-calculated, using the new values for the load temperatures and/or
efficiencies. The calibration parameters in the raw data are replaced by tho
set with OVERRIDE until reset by OVERRIDE /RESET
Keyword TEM[PLOAD]: changes the temperatures of the cold load (value1)
and of the hot load (value2).
Keyword EFF[ICIENCY]: changes the forward efficiency (value1) and the
beam efficiency (value2).
Keyword GAIN[IMAGE]: changes the the suppression of the image band
(value1).
OVERRIDE /FEBE ifb|all This is to specify the frontend-backend combination
(ifb) for which the calibration parameters are to
be changed. Keyword "all" changes the parameters
for all frontend-backend combinations in the
current scan.
OVERRIDE /RESET Resets all calibration parameters to default (raw da
values). A new calibration has to be read in order t
activate the change.
EXAMPLES:
OVERRIDE TEMPLOAD * 300 /FEBE all Changes the hot load temperature for all
frontends to 300 K. All cold load
temperatures are left unchanged.
OVERRIDE EFFICIENCY 0.95 Changes the forward efficiency of
frontend 1 to 0.95. All beam
efficiencies are unchanged.
OVERRIDE GAINIMAGE 0.1 /FEBE 1 Changes the sideband suppression for the
receiver in frontend-backend combination
number 1 to 0.1.
OVERRIDE TAMBPID -7.5 712 Changes ambient outside temperature and
pressure to -7.5 deg C and 712 hPa.